Specially designed for containment of DNA carry over and neutralization prevents cross contamination down from previous amplified DNA fragments.
Features
- Partially recirculating downflow cabinet with air being filtered by a HEPA (H14) filter with 99.995% overall collection efficiency on 0.1 – 0.2 m particles (MPPS), before entering the work area at a speed of 0.45 to 0.55 m/sec; this ensures a Class 100 (FED STD 209E) – ISO 5 (ISO 14644-1 Standard) environment throughout the work area.
- Two configurations are available: INWARD air barrier and OUTWARD air barrier, that can be easily chosen by a simple mechanical operation.
- INWARD air barrier. An air barrier is pulled through the front opening and recirculated with the downflow air by a motor blower. 70% of the air is returned to the work area through the main HEPA (H14) filter and 30% is exhausted into the environment through an air filter with gravimetric efficiency of 99% ensuring excellent product protection as well as an outstanding containment.
- Removable perforated work surface and back wall of the work chamber made of AISI 304 stainless steel
- Front panels in tempered glass
- Hinged front window
- Centrifugal Motor blower
- Soft-touch keys on the control panel provide control of fan and fluorescent lamps lighting
- Elapsed time-meter for HEPA filter (Option)
- Tubular Stand (Option)
- UV lamp.
Technical Specifications & Ordering Informations | |
Model | LXMPCR-2.5 |
Work Chamber Dimensions (W x D x H) mm | 745 x 435 x 500 mm |
Overall Dimensions (W x D x H) mm | 820 x 580 x 790 mm |
Front Opening Height (mm) | 180 |
Power Supply | 220/240 V – 50 Hz |
Power W | 200 |
Noise Level | dB(A) ≤ 65 db at 1.5 mtr. distance in open field condition |
Lighting Lux | 1200 |
Tubular Stand (option) | MS tubular stand duly powder coated (option) |
Options
- Elapsed time-meter for HEPA filter.
- MS tubular stand duly powder coated.
- TFT Touch Screen Display Controller with air velocity sensor.